Comparison of the cyclic fatigue resistance of hyflex EDM, vortex blue, protaper gold, and onecurve nickel-Titanium instruments
Tarih
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Erişim Hakkı
Özet
Objective: This in vitro study aimed to compare the cyclic fatigue resistance of HyFlex EDM (HEDM), Vortex Blue (VB), ProTaper Gold (PG), and OneCurve (OC) nickel-titanium (NiTi) instruments. Materials and Methods: About 12 HEDM (25/.similar to), 12 VB (25/.06), 12 PG (25/.08), and 12 degrees C (25/.06) instruments were included in this study. All the instruments were tested with a 60 degrees angle of curvature and a 3-mm radius of curvature. All the instruments were utilized until fracture occurred, and then the number of cycles to failure (NCF) was calculated. The data were analyzed statistically using Kruskal-Wallis H and Mann-Whitney U-tests. The statistical significance level was set at P < 0.05. Results: HEDM instruments had the highest cyclic fatigue resistance among all the other instruments (P < 0.05). The OC instruments had a significantly higher fatigue resistance than the PG and VB instruments (P < 0.05); however, there was no significant difference between PG and VB instruments in the NCF (P > 0.05). Conclusion: Within the limitations of this in vitro study, it was found that the cyclic fatigue resistance was higher for the HEDM instruments than for the VB, PG, and OC instruments.
Açıklama
PubMed: 31929205